Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 13 de 13
Filter
1.
J Cancer Res Ther ; 2019 Oct; 15(5): 1011-1017
Article | IMSEAR | ID: sea-213470

ABSTRACT

Context: Electronic portal imaging devices (EPIDs) could potentially be useful for patient setup verification and are also increasingly used for dosimetric verification. The accuracy of EPID for dose verification is dependent on the dose-response characteristics, and without a comprehensive evaluation of dose-response characteristics, EPIDs should not be used clinically. Aims: A scatter correction method is presented which is based on experimental data of a two-dimensional (2D) ion chamber array. An accurate algorithm for 2D dose reconstruction at midplane using portal images for in vivo dose verification has been developed. Subjects and Methods: The procedure of scatter correction and dose reconstruction was based on the application of several corrections for beam attenuation, and off-axis factors, measured using a 2D ion chamber array. 2D dose was reconstructed in slab phantom, OCTAVIUS 4D system, and patient, by back projection of transit dose map at EPID-sensitive layer using percentage depth dose data and inverse square. Verification of the developed algorithm was performed by comparing dose values reconstructed in OCTAVIUS 4D system and with that provided by a treatment planning system. Results: The gamma analysis for dose planes within the OCTAVIUS 4D system showed 98% ±1% passing rate, using a 3%/3 mm pass criteria. Applying the algorithm for dose reconstruction in patient pelvic plans showed gamma passing rate of 96% ±2% using the same pass criteria. Conclusions: An accurate empirical algorithm for 2D patient dose reconstruction has been developed. The algorithm was applied to phantom and patient data sets and is able to calculate dose in the midplane. Results indicate that the EPID dose reconstruction algorithm presented in this work is suitable for clinical implementation

2.
Clinics ; 74: e835, 2019. tab, graf
Article in English | LILACS | ID: biblio-1011902

ABSTRACT

The efficacy and toxicity of radionuclide therapy are believed to be directly related to the radiation doses received by target tissues; however, nuclear medicine therapy continues to be based primarily on the administration of empirical activities to patients and less frequently on the use of internal dosimetry for individual therapeutic planning. This review aimed to critically describe the techniques and clinical evidence of dosimetry as a tool for therapeutic planning and the main limitations to its implementation in clinical practice. The present article is a nonsystematic review of voxel-based dosimetry. Clinical evidence pointing to a correlation between the radiation dose and therapeutic response in various diseases, such as thyroid carcinoma, neuroendocrine tumors and prostate cancer, is reviewed. Its limitations include technical aspects related to image acquisition and processing and the lack of randomized clinical trials demonstrating the impact of dosimetry on patient therapy. A more widespread use of dosimetry in therapeutic planning involves the development of user-friendly dosimetric protocols and confirmation that dose estimation implies good efficacy and low treatment-related toxicity.


Subject(s)
Humans , Radioisotopes/administration & dosage , Radiometry/methods , Radiotherapy/methods , Radionuclide Imaging/methods , Time Factors , Tomography, Emission-Computed, Single-Photon/methods , Positron-Emission Tomography/methods
3.
Chinese Journal of Radiation Oncology ; (6): 833-837, 2017.
Article in Chinese | WPRIM | ID: wpr-620207

ABSTRACT

In vivo dosimetry (IVD) is currently the most direct and effective means of quality assurance.The electronic portal imaging device (EPID) has been widely used for IVD verification owing to its favorable dosimetric properties.In recent years,an increasing number of EPID-based IVD studies have emerged around the world.The purpose of this paper is to give an overview of the present progress in EPID-based IVD studies,and to provide a reference for the subsequent application of EPID in IVD.

4.
Chinese Journal of Radiological Medicine and Protection ; (12): 298-301, 2017.
Article in Chinese | WPRIM | ID: wpr-512180

ABSTRACT

Objective To study the difference of the constructed doses between electronic portal imaging device (EPID) and dynalogs files of linac for in vivo phantom dosimetry.Methods Twelve pelvic patients treated with volumetric modulated arc therapy (VMAT) plans were selected and the information of each plan was copied to theCheese phantom to recalculate the doses before delivered on Varian RapidArc Linac.TheCheese phantom was placed on the isocenter and the electronic portal image (EPI) formed by the EPID was sent to EPIgray software to reconstruct the actual delivered doses.Meanwhile,dynalogs files were respectively imported to the Mobius software to reconstruct the actual delivered doses too.The point dose in the center of each VMAT plan (the center of the effective sensitive volume of ionization chamber) was measured by the A1SL ionization chamber.At the same time,the dose of sensitive volume of ionization chamber from treatment planning systcm (TPS) was recorded.Results The relative deviation between the dose from TPS and the measurement results by the ionization chamber was 1.13%.The difference between the reconstructed doses of EPID-based or the dynalogs file-based with the measurement results by the ionization chamber was not statistically significant (P > 0.05).Conclusions The two methods of dose reconstruction can provide reference for in vivo dosimetry of VMAT.

5.
Radiation Oncology Journal ; : 121-128, 2017.
Article in English | WPRIM | ID: wpr-44441

ABSTRACT

PURPOSE: To report the results of a correlation analysis of skin dose assessed by in vivo dosimetry and the incidence of acute toxicity. This is a phase 2 trial evaluating the feasibility of intraoperative radiotherapy (IORT) as a boost for breast cancer patients. MATERIALS AND METHODS: Eligible patients were treated with IORT of 20 Gy followed by whole breast irradiation (WBI) of 46 Gy. A total of 55 patients with a minimum follow-up of 1 month after WBI were evaluated. Optically stimulated luminescence dosimeter (OSLD) detected radiation dose delivered to the skin during IORT. Acute toxicity was recorded according to the Common Terminology Criteria for Adverse Events v4.0. Clinical parameters were correlated with seroma formation and maximum skin dose. RESULTS: Median follow-up after IORT was 25.9 weeks (range, 12.7 to 50.3 weeks). Prior to WBI, only one patient developed acute toxicity. Following WBI, 30 patients experienced grade 1 skin toxicity and three patients had grade 2 skin toxicity. Skin dose during IORT exceeded 5 Gy in two patients: with grade 2 complications around the surgical scar in one patient who received 8.42 Gy. Breast volume on preoperative images (p = 0.001), ratio of applicator diameter and breast volume (p = 0.002), and distance between skin and tumor (p = 0.003) showed significant correlations with maximum skin dose. CONCLUSIONS: IORT as a boost was well-tolerated among Korean women without severe acute complication. In vivo dosimetry with OSLD can help ensure safe delivery of IORT as a boost.


Subject(s)
Female , Humans , Breast Neoplasms , Breast , Cicatrix , Follow-Up Studies , Incidence , Luminescence , Mastectomy, Segmental , Radiotherapy , Seroma , Skin
6.
Biomedical Imaging and Intervention Journal ; : 1-12, 2009.
Article in English | WPRIM | ID: wpr-625887

ABSTRACT

The authors have studied the feasibility of using three new high-sensitivity radiochromic devices in measuring the doses to peripheral points outside the primary megavoltage photon beams. The three devices were GAFCHROMIC® EBT film, prototype Low Dose (LD) Film, and prototype LD Card. The authors performed point dosimetry using these three devices in water-equivalent solid phantoms at x = 3,5,8,10, and 15 cm from the edge of 6 MV and 15 MV photon beams of 10x10 cm2, and at depths of 0, 0.5 cm, and depth of maximum dose. A full sheet of EBT film was exposed with 5000 MU. The prototype LD film pieces were 1.5x2 cm2 in size. Some LD films were provided in the form of a card in 1.8x5 cm2 holding an active film in 1.8x2 cm2. These are referred to as “LD dosimeter cards”. The small LD films and cards were exposed with 500 MU. For each scanned film, a 6 mm circular area centered at the measurement point was sampled and the mean pixel value was obtained. The calibration curves were established from the calibration data for each combination of film/cards and densitometer/scanner. The doses at the peripheral points determined from the films were compared with those obtained using ion chamber at respective locations in a water phantom and general agreements were found. It is feasible to accurately measure peripheral doses of megavoltage photon beams using the new high-sensitivity radiochromic devices. This near real-time and inexpensive method can be applied in a clinical setting for dose measurements to critical organs and sensitive patient implant devices.

7.
Korean Journal of Medical Physics ; : 48-54, 2007.
Article in Korean | WPRIM | ID: wpr-27792

ABSTRACT

This study is to develope a phantom for MOSFET (Metal Oxide Semiconductors Field Effect Transistors) dosimetry and compare the dosimetric properties of standard MOSFET and microMOSFET with the phantom. In this study, the developed phantom have two shape: one is the shape of semi-sphere with 10 cm diameters and the other one is the flat slab of 30 cm x 30 cm with 1 cm thickness. The slab phantom was used for calibration and characterization measurements of reproducibility, linearity and dose rate dependency. The semi-sphere phantom was used for angular and directional dependence on the types of MOSFETs. The measurements were conducted under 10 x 10 cm2 fields at 100 cm SSD with 6 MV photon of Clinac (21EX, Varian, USA). For calibration and reproducibility, five standard MOSFETs and microMOSFETs were repeatedly irradiated by 200 cGy five times. The average calibration factor was a range of 1.09+/-0.01~1.12+/-0.02 mV/cGy for standard MOSFETs and 2.81+/-0.03~2.85+/-0.04 mV/cGy for microMOSFETs. The response of reproducibility in the two types of MOSFETs was found to be maximum 2% variation. Dose linearity was evaluated in the range of 5 to 600 cGy and showed good linear response with R2 value of 0.997 and 0.999. The dose rate dependence of standard MOSFET and microMOSFET was within 1% for 200 cGy from 100 to 600 MU/min. For linearity, reproducibility and calibration factor, two types of MOSFETs showed similar results. On the other hand, the standard MOSFET and microMOSFET were found to be remarkable difference in angular and directional dependence. The measured angular dependence of standard MOSFET and microMOSFET was also found to be the variation of 13%, 10% and standard deviation of +/-4.4%, +/-2.1%. The directional dependence was found to be the variation of 5%, 2% and standard deviation of +/-2.1%, +/-1.5%. Therefore, dose verification of radiation therapy used multidirectional X-ray beam treatments allows for better the use of microMOSFET which has a reduced angular and directional dependence than that of standard MOSFET.


Subject(s)
Calibration , Hand , Semiconductors , Silver Sulfadiazine
8.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 155-163, 2004.
Article in Korean | WPRIM | ID: wpr-52744

ABSTRACT

PURPOSE: To confirm the reproducibility of in vivo transmission dosimetry system and the accuracy of the algorithms for the estimation of transmission dose in head and neck radiation therapy patients. MATERIALS AND MEHTODS: From September 5 to 18, 2001, transmission dose measurements were performed when radiotherapy was given to brain or head and neck cancer patients. The data of 35 patients who were treated more than three times and whose central axis of the beam was not blocked were analyzed in this study. To confirm the reproducibility of this system, transmission dose was measured before daily treatment and then repetitively every hour during the treatment time, with a field size of 10x10 cm2 and a delivery of 100 MU. The accuracy of the transmission dose calculation algorithms was confirmed by comparing estimated dose with measured dose. To accurately estimate transmission dose, tissue inhomogeneity correction was done. RESULTS: The measurement variations during a day were within +/-0.5% and the daily variations in the checked period were within +/-1.0%, which were acceptable for system reproducibility. The mean errors between estimated and measured doses were within +/-5.0% in patients treated to the brain, +/-2.5% in head, and +/-5.0% in neck. CONCLUSION: The results of this study confirmed the reproducibility of our system and its usefulness and accuracy for daily treatment. We also found that tissue inhomogeneity correction was necessary for the accurate estimation of transmission dose in patients treated to the head and neck.


Subject(s)
Humans , Axis, Cervical Vertebra , Brain , Brain Neoplasms , Head and Neck Neoplasms , Head , Neck , Radiotherapy
9.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 276-282, 2003.
Article in Korean | WPRIM | ID: wpr-126378

ABSTRACT

PURPOSE: Although high-dose-rate intracavitary radiotherapy (HDR ICR) has been used in the treatment of cervical cancer, the potential for increased risk of late complication, most commonly in the rectum, is a major concern. We have previously reported on 136 patients treated with HDR brachytherapy between 1995 and 1999. The purpose of this study is to upgrade the previous data and confirm the correlation between late rectal complication and rectal dose in cervix cancer patients treated with HDR ICR. MATERIALS AND METHODS: A retrospective analysis was performed for 222 patients with cervix cancer who were treated for curative intent with external beam radiotherapy (EBRT) and HDR ICR from July 1995 to December 2001. The median dose of EBRT was 50.4 (30.6~56.4) Gy with a daily fraction size 1.8 Gy. A total of six fractions of HDR ICR were given twice weekly with fraction size of 4 (3~5.5) Gy to A point by Iridium-192 source. The rectal dose was calculated at the rectal reference point using the barium contrast criteria. in vivo measurement of the rectal dose was performed with thermoluminescent dosimeter (TLD) during HDR ICR. The median follow-up period was 39 months, ranging from 6 to 90 months. RESULTS: Twenty-one patients (9.5%) experienced late rectal bleeding, from 3 to 44 months (median, 13 months) after the completion of RT. The calculated rectal doses were not different between the patients with rectal bleeding and those without, but the measured rectal doses were higher in the complicated patients. The differences of the measured ICR rectal fractional dose, ICR total rectal dose, and total rectal biologically equivalent dose (BED) were statistically significant. When the measured ICR total rectal dose was beyond 16 Gy, when the ratio of the measured rectal dose to A point dose was beyond 70%, or when the measured rectal BED was over 110 Gy3, a high possibility of late rectal complication was found. CONCLUSION: Late rectal complication was closely correlated with measured rectal dose by in vivo dosimetry using TLD during HDR ICR. If data from in vivo dosimetry shows any possibility of rectal bleeding, efforts should be made to reduce the rectal dose.


Subject(s)
Humans , Barium , Brachytherapy , Follow-Up Studies , Hemorrhage , Radiotherapy , Rectum , Retrospective Studies , Uterine Cervical Neoplasms
10.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 274-282, 2002.
Article in Korean | WPRIM | ID: wpr-81188

ABSTRACT

PURPOSE: Measurement of transmission dose is useful for in vivo dosimetry. In this study, the algorithm for estimating the transmission dose for open radiation fields was modified for application to partially blocked radiation fields. MATERIALS AND METHODS: The beam data was measured with a flat solid phantom with various blocked fields. A new correction algorithm for partially blocked radiation field was developed from the measured data. This algorithm was tested in some settings simulating clinical treatment with an irregular field shape. RESULTS: The correction algorithm for the beam block could accurately reflect the effect of the beam block, with an error within +/-1.0%, with both square fields and irregularly shaped fields. CONCLUSION: This algorithm can accurately estimate the transmission dose in most radiation treatment settings, including irregularly shaped field.

11.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 186-192, 2002.
Article in Korean | WPRIM | ID: wpr-190468

ABSTRACT

PURPOSE: Measurement of transmission dose is useful for in vivo dosimetry. In this study, previous algorithm for estimation of transmission dose was modified for use in cases with tissue deficit. MATERIALS AND METHODS: The beam data was measured with flat solid phantom in various conditions of tissue deficit. New algorithm for correction of transmission dose for tissue deficit was developed by physical reasoning. The algorithm was tested in experimental settings with irregular contours mimicking breast cancer patients using multiple sheets of solid phantoms. RESULTS: The correction algorithm for tissue deficit could accurately reflect the effect of tissue deficit with errors within +-1.0% in most situations and within +-3.0% in experimental settings with irregular contours mimicking breast cancer treatment set-up. CONCLUSION: Developed algorithm could accurately reflect the effect of tissue deficit and irregularly shaped body contour on transmission dosimetry.


Subject(s)
Humans , Breast Neoplasms
12.
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 293-299, 2001.
Article in Korean | WPRIM | ID: wpr-202268

ABSTRACT

PURPOSE: LiF TLD has a problem to be used in vivo dosimetry because of the toxic property of LiF. The aim of this study is to develop new dosimeter with LiF TLD to be used in vivo dosimetry. MATERIALS AND METHODS: We designed and manufactured the teflon box(here after TLD holder) to put TLD in. The external size of TLD holder is 4x4x1 mm3. To estimate the effect of TLD holder on TLD response for radiation, the linearity of TLD response to nominal dose were measured for TLD in TLD holder. Measurement were performed in the 10 MV x-ray beam with LiF TLD using a solid water phantom at SSD of 100 cm. Percent Depth Dose (PDD) and Tissue-Maximum Ratio (TMR) with varying phantom thickness on TLD were measured to find the effect of TLD holder on the dose coefficient used for dose calculation in radiation therapy. RESULTS: The linearity of response of TLD in TLD holder to the nominal dose was improved than TLD only used as dosimeter. And in various measurement conditions, it makes a marginnal difference between TLD in TLD holder and TLD only in their responses. CONCLUSION: It was proven that the TLD in TLD holder as a new dosimetry could be used in vivo dosimetry.


Subject(s)
Polytetrafluoroethylene , Silver Sulfadiazine , Water
13.
Journal of the Korean Society for Therapeutic Radiology ; : 219-224, 1994.
Article in Korean | WPRIM | ID: wpr-125378

ABSTRACT

This paper describes the basic date measurements for total body irradiation with 6 Mv photon beam including compensators designs. The technique uses bilateral opposing field with tissue compensators for the head, neck, lungs, and legs from the hip to toes. In vivo dosimetry was carried out for determining absorbed dose at various regions in 7 patients using diode detectors (MULTIDOSE, Model 9310, MULTIDATA Co., USA). As a results, the dose uniformity of+/-3.5%(generally, within+/-10%) can be achieved with our total body irradiation technique.


Subject(s)
Humans , Head , Hip , Leg , Lung , Neck , Toes , Whole-Body Irradiation
SELECTION OF CITATIONS
SEARCH DETAIL